Endress-Hauser Memosens CCS51E là Cảm biến đo Clo dành cho nước uống, nước hồ bơi và nước công nghiệp, ...
- Tài liệu kỹ thuật của ở đây.
- Tài liệu hướng dẫn vận hành của ở đây.
Overview
Specs at a glance
Measuring range
Trace: 0 to 5 mg/l HOCl
Standard: 0 to 20 mg/l HOCl
High: 0 to 200 mg/l HOCl
Process temperature
0 to 55 °C (32 to 130 °F), non-freezing
Process pressure
Max. 1 bar (max. 14.5 psi)
Measuring method
Closed, membrane covered measuring cell
Reduction of free chlorine at the cathode
Field of application
Memosens CCS51E ensures precise and consistent disinfection monitoring in drinking water, pool and process water and in utilities. This sensor for determining free chlorine features extremely fast response times even at low water volumes. This guarantees efficient process control and safe processes. Thanks to Memosens 2.0 technology, CCS51E combines maximum process and data integrity with simple operation. This provides the perfect basis for predictive maintenance and IIoT connectivity.
The free chlorine sensor Memosens CCS51E is suitable for a wide range of water qualities:
Drinking water: to secure disinfection
Utilities: to detect free chlorine and avoid any damage
Process water: for hygienic and safe packaging and bottling of products in food processing
Pool water: to dose disinfectant efficiently
Benefits
The low-maintenance, amperometric sensor reduces the cost of ownership of the measuring point, particularly compared to colorimetric measuring systems.
Increased process up-time: pre-calibration of the sensor allows for real plug & play and faster polarization time thanks to Memosens 2.0 technology.
Fast response time (t90 < 25 s) provides an accurate process view and enables prompt reaction to process changes as well as efficient process control.
High process safety and efficient dosing: precise and long-term stable measurement ensures consistent process monitoring and permits individually adapted disinfectant dosing.
Suitable sensor versions for every measurement range: From trace measurement up to free chlorine concentrations of 200 mg/l.
Connection to the Liquiline multiparameter transmitter facilitates easy combination with other parameters of liquid analysis such as pH and ORP.
Specifications
Measuring / Application task
- Disinfection
Measuring principle
Free chlorine
Application
Ensuring reliable disinfection in drinking water
Process water
Dosing disinfectant efficiently in pool water
Detect the absence or presence of free chlorine in Utilities
Guarantee food safety and provide hygienic packaging and bottling
Characteristic
Amperometric measurement of dissolved free chlorine
Measuring range
Trace: 0 to 5 mg/l HOCl
Standard: 0 to 20 mg/l HOCl
High: 0 to 200 mg/l HOCl
Measuring method
Closed, membrane covered measuring cell
Reduction of free chlorine at the cathode
Design
Closed amperometric 2-electrode measuring cell with PVDF membrane
Material
Sensor shaft: POM
Membrane: PVDF
Membrane cap: PVDF
Sealing ring: FKM
Dimension
Diameter: 25 mm (0.98 inch)
Length: 161 mm (6.34 inch)
Process temperature
0 to 55 °C (32 to 130 °F), non-freezing
Process pressure
Max. 1 bar (max. 14.5 psi)
Temperature sensor
10k NTC integrated (Memosens)
Connection
Inductive, digital connection head with Memosens
Spare parts & accessories
8 spare parts & accessories
Maintenance kits CCV05
Order code CCV05-
100 pcs. of DPD tablet Nr. 1
Order code 50035461
Kit adapter CCS5x(D/E) for CCA250
Order code 71372025
Kit adapter CCS5x(D/E) for CYA112
Order code 71372026
Kit adapter CCS5x(D/E) for CYA27+CCA151
Order code 71372027
Kit adapter CCS5x(D/E) for CYA27-xxxB
Order code 71554885
Zero-point gel COY8
Order codeCOY8-
photometer M+N PF-3, disinfection
Order code 71257946
Related products
6 products & bundles
Flow assembly for multi-parameter measurements Flowfit CYA27
Modular assembly for drinking and process water applications
Process temperature
0 to 60 °C (32 to 140°F)
Process pressure
0 to 4 bar (0 to 58 psi) relative
1-/2-channel transmitter
Liquiline CM442
Expandable multiparameter field device for all industries
Input
1 to 2x Memosens digital input
Output / communication
2 to 8x 0/4 to 20 mA current outputs
Alarmrelay, 2x relay
Ingress protection
IP67
1-/2-channel transmitter
Liquiline CM442R
Compact multiparameter DIN-rail and cabinet device for all industries
Input
1 to 2x Memosens digital input
2x 0/4 to 20mA Input optional
2x Digital input optional
Output / communication
2 to 8x 0/4 to 20 mA current outputs
Alarmrelay, 2x relay, ProfibusDP, Modbus RS485,
Modbus TCP, Ethernet
Ingress protection
Transmitter: IP20
Optional Display: IP66
Digital pH sensor
Memosens CPS31E
Memosens 2.0 glass electrode for application in drinking water and swimming pools
Measuring range
pH 1 to 12
Process temperature
–15 to 80 °C (5 to 176 °F)
Process pressure
0.8 to 4 bar (11.6 to 58 psi) (absolute)
Free chlorine process water panel
Ready-to-use panel bundle for reliable and precise monitoring of disinfection processes
Multiparameter drinking water panel
Ready-to-use panel bundle for measurement of the most important drinking water parameters pH, conductivity, turbidity and free chlorine